Rapid determination of Solid Fat Content (SFC) in Oils and Fats using Spin Track NMR Analyzer. Printable version (128 KB)
Solid Fat Content (SFC) is generally accepted analysis of fats and oils in the food industry. The traditional extraction methods for SFC determination are slow, irreproducible and require additional chemicals. Direct measurements of SFC by NMR (nuclear magnetic resonance) provides quick and accurate determination of SFC value. NMR as method of fats and oils analysis is included in the following international standards:
Solid Fat Content determination by NMR is based on direct ratio measuring between the solid and liquid parts of the sample. After excitation of the sample by powerful 90o RF pulse FID (Free Induction Decay) is detected. FID is the signal accompanies relaxation process of hydrogen proton magnetic spins back to equilibrium state after disturbing. FID contains contributions from both solid and liquid parts. The protons of liquid content are more rapid in comparsion with protons of solid. It provides more fast signal decay from solids than from liquids. Thus, these contributions can be distinguished.
Fig. 1. Principle of SFC determination using FID The SFC value is determinated by taking two measuring points on FID. FID amplitude at point S corresponding to total solids plus liquids and at point L corresponding liquids only The specific ratio can be found using equation shown on Fig. 1. This ratio is considered as the SFC value. An F-factor (f) allows prediction of FID amplitude value immediately after RF pulse as it is impossible to obtain initial NMR signal amplitude direclty during so-called dead-time. Dead-time is a short time (< 10 us) period caused by the decay of powerful RF pulse in resonance circuit of NMR sensor. F-factor increases the measurements accuracy and is determined during calibration procedure. Portable NMR spectrometer Spin Track from Resonance System Ltd. is ideal instrument for SFC measurements because of fine technical characteristics which make measurements very reproducible and accurate. Small weight and mobility allow its usage in production routine measurements and in advanced laboratories for complicated investigations. Spin Track fulfills the requirements of international standards like AOCS Cd 16b-93, AOCS Cd 16-81 revised in 2000, ISO 8292, IUPAC 2.150. The volume of the sample is 1-3 ml. SFC measurements cycle for 6 point melting curve using special container for samples and tempering:
The procedure of calibration can be performed using the set of original calibration samples developed by Resonance Systems Ltd. |
|